Categories
Price: | US $ 2135/Piece |
---|---|
Trade Terms: | FOB,CIF,CPT |
Min Order: | 1/Piece |
Pay Type: | L/C,T/T,Western Union |
n Master-Q series deionized pure water system(Tap water inlet) | |||||
Model | Deionized water purification system | ||||
Master-Q | Master-Q UT | Master-Q plus | Master-Q plus UT | ||
Flow procedure | PF+KDF+AC+RO+AC+DI | PF+KDF+AC+RO + UV+AC + DI +TF | PF+KDF+AC+RO+AC+DI | PF+KDF+AC+RO + UV+AC + DI +TF | |
Application | Ware washing, Agricultural expe, General biological expe, Aquatic products feeding, Inlet water for Ultra pure water machine, Inlet water for sterilizer/T&H chamber, Buffer disposing, Aseptic drinking water, Physical and chemical analysis, Fine chemistry industry, Inlet water for Ultra pure water machine, GC/HPLC, Pharmacy research, Medicine examining | ||||
DI water quality | Resistivity | 16-18.2MΩ.cm | |||
Heavy metal ions | 0.1ppb | ||||
TOC | 30 ppb | ||||
Bacteria | - | 1CFU/ml | - | 1CFU/ml | |
Particle(>0.22μm) | - | 1/ml | - | 1/ml | |
RO water quality | < TDS(ppm, mg/l) of tap water×5% | ||||
Technical specification |
Output(25 degrees celsius) | 15 Liters/hour* | 30 Liters/hour | ||
Flow rate | 2 Lliters/min(with pressure tank) | ||||
Pure water outlet | RO water, deionized water | ||||
Dimension/weight | Length×Width×Height:50×36×54cm / 20-30Kg | ||||
Electric requirements | AC110-220V,50/60Hz,120W | ||||
Control system |
Mode display | Power on, program, inlet rinse, producing, full, circle, regular outlet, disinfection, consumables replacing remind | |||
Safety | Low pressure and full water alarm, password, auto-reset, outlet forbidden when alarm or disinfection | ||||
System monitor | Display quality of feed water, RO water and ultrapure water, temperature, used and left time of consumables | ||||
Feed water requirements |
Tap water, inlet TDS<200 ppm,5-45 degrees celsius,1.0-4.0kgf/cm 2 (if inlet TDS>200ppm, extra pretreatment is recommended) |
||||
Purification system |
Pretreatment unit | 5μm spun fiber filter×1+Kinetic degradation fluxion filter×1+Granular active carbon filter×1 | |||
RO unit | 100 GPD RO membrane×1 ( Plus model: 100 GPD RO membrane×2) | ||||
Subsequent unit |
Post active carbon filter×1+Mixed bed resin cartridge×3{Plus model: Mixed bed resin cartridge×4} UT model:+254nm wavelength UV lamp cartridge×1+(0.45+0.1) μm terminal filter×1 |
||||
Standard configuration | Main body(including:1 set cartridge)+4 gallon tank+ accessory box |
Prod Model: | Master-Q UT |
---|---|
Markets: | North America,South America,Eastern Europe,Southeast Asia,Africa,Oceania,Mid East,Eastern Asia,Western Europe |
Flow Rate: | 15 Liter/Hour |
Resistivity: | 13-17.5 Megohm-Cm |
Heavy Metal Ion(Ppb): | <0.1ppb |
Toc: | <30ppb |
Bacteria: | <1cfu/Ml |
Granular(>0.22um): | <1/Ml |
RO Water TDS(Ppm): | <10* |
Outlet: | RO Water,Di Water |
Dimension/Weight: | Lxwxh:50X36X54cm/ About22kg |
Power: | 100-250V,50-60Hz/120W |
Type: | Separation Type |