| Apparatus: | Coal and Oils Oxygen Bomb Calorimeter |
| Application: | Fuel Oxygen Bomb Calorimeter |
| Function: | Coal Oxygen Bomb Calorimeter |
| Device: | Laboratory Oxygen Bomb Calorimeter |
| Heat Capacity: | 14000-15000j/K |
| Temperature Measurement Range: | 10 -35c |
| Temperature Resolution: | 0.001 C |
| Ambient Humidity: | Less Than 85% |
| Shipping Size: | 71*56*50cm;46*37*26cm |
| Warranty: | 1 Year |
| 1 | Heat capacity | 14000 J/K~15000 J/K |
| 2 | Measuring range | (10~35)°C |
| 3 | Resolution | 0.001 K |
| 4 | Measurement accuracy | ± 60J/g |
| 5 | Repeatability error | ≤0.2% (Grade C) |
| 6 | Pressure endurance of oxygen bomb | 20 MPa |
| 7 | Ambient temperature | 10 °C~28 °C |
| 8 | Relative humidity | ≤85% |
| 9 | Power supply | AC 220V±5%, 50 Hz |
| 10 | Dimension | 600mm×480mm×460mm |
| Parameters | GDY-1A | GDY-1A+ | GDY-1C |
| Test mode | Manual | Semi-auto | Automatic |
| Control model | In-build computer | Micro-computer | Computer |
| Packing Size | 72*59*51 47*47*28 | 71*56*50 46*37*26 | 69*55*48 45*37*25 58*52*84 |
| Gross Weight | 40kg | 33kg | 63kg |
| No. | Problem | Cause | Solution |
| 1 | Indication light is off. | 1.No power supply. 2.The fuse is damaged. | 1.Connect the power supply 2.Replace the fuse. |
| 2 | Can not be ignited. | 1.The ignition electric pole is poor connected with oxygen bomb. 2.The ignition wire is short circuited. | 1.Make sure the electric pole and oxygen bomb well connected. 2.Install the ignition wire again. |
| 3 | Stirrer doesn't work | 1.The stirring motor is blocked. 2.The stirring motor wire is disconnected. | 1.Get rid of the blocks. 2.Connect the wire again. |
| 4 | Temperature doesn't decrease | 1.Water temperature in outer bucket is much lower than ambient temperature. 2.The sample mass is too less. | 1.Make determination again after regulating the outer bucket temperature. 2. 2.Use more sample to rise the temperature 2 K~3K. |
| 5 | Clock stops | 1. the clock has no power | 1.Replace battery 2.Manually input time to continue determination. |