Product Specifications
| Prod Model: |
TPS-GS-400 |
| Markets: |
North America,South America,Eastern Europe,Southeast Asia,Africa,Oceania,Mid East,Eastern Asia,Western Europe |
Product Description
GS 400 auto chemistry analyzer
System function
Throughput: Up to 450 tests/Hour(not including ISE tests)
Test Method: End point, Fixed time, Kinetic
Test Sequence: Discrete, Random access, STAT can be inserted at any time. Test sequence can be
sorted by samples or items
Spectrometer: Enclosed array of spectrometer and photodiodes, maintenance free.
Cuvette washing: Auto wash the cuvettes through 6 segments washing pipe and continuous use
Them, and each cuvette wash 12 times before use it again.
Reaction detect: Monitor 81 cuvettes's reaction at the same time
Photometric process: The interval time between two consecutive photometric points are exactly the
same.
Pipeline system: Pressure auto released pumps and double filters to protect pumps and valves
Gearing: Plastic gear leads lower noise and accuracy of positioning of moving parts
Environmental resistance: Resistant form dust and water vapor
Sample/Reagent/Stirrer system
Sample volume: 2μL~30μL, step by 0.1μL
Sample disc: 60 positions for primary tubes and secondary cups
Sample probe: Inner and outer walls are highly polished leads carry-over <0.1%, level detection
and vertical and horizontal collision protection
Sample tube: Both primary tubes and secondary cups are suitable
Reagent volume: 20μL~300μL, step by 1μL
Reagent disc: Provide 80 reagent bottle positions
Reagent probe: Inner and outer walls are highly polished leads carry-over <0.1%, level detection
and vertical and horizontal collision protection
Reagent bottle: Particular reagent bottle without dead volume
Reagent Cooling: Continuous cooling system according to the principle of peltier
Stirrer: Paddle stirrer coated with Teflon, stir immediately after the sample was added in.
Water consume: <6L/hour
Carry over: Optimized item test sequence and reinforced washing process, lower the carry over
and avoid probe being clotted
Reaction system
Cuvette: Permanent UV free glass cuvette
Reaction disc: Discal reaction disc with 81 disposable cuvettes
Reaction volume: 150μL~600μL
Heater: Solid and sealed heater leads temperature raise rapidly and steady and maintenance free
Temperature: 37º C± 0.1º C
Optical system
Light source: 12V 20W, halogen lamp with lifetime more than 2000 hours
Spectrometer: Post spectrometer with arrayed and sealed filters, maintenance free
Wavelength: 340nm\405nm\450nm\510nm\546nm\578nm\630nm\670nm
Absorbance range: 0~6
Calibration/QC
Calibration type: Single point linear\two points linear\multi-points linear\Logit-4P\Logit-5P\
Spline\Exponential 5P\Polynomial 5P, auto choose the best fittest calibration
curve.
QC type: Real time QC\daily QC\day to day QC
QC chart: L-J\Cumulative sum check\Twin Plot
QC rule: Customized QC rule in order to adapt different item's requirements
Operational system
Operational system: VISTA\Windows XP\Windows2000
Memory: Auto save and recover the results
Report format: Customized report format
Status Check: Real time check the working status of the analyzer
Power/ Dimension
Voltage: AC 85V~264V, 55Hz± 8Hz, Auto adapt according to the input voltage
Power Consume: ≤ 550VA
Dimension: 980mm× 712mm× 1100mm(L× W× H)
Net weight: 160Kg
Shanghai Toposun Industrial Co., Ltd. is an internationally leading laboratory and process technology provider with areas of core competence in biotechnology and medical. The products of our company are widely applied in microelectronics, optics, precise machinery, medical and health, bio-pharmaceuticals, food and drinks, ultra-purification smelting, aviation and space flight, and other industrial aspects and the research in colleges, universities and institutes and the asepsis sickrooms in hospitals, etc.
Inspection instrument:
Dust particle counter, dust particle multi-point inspection system, sampling device, environment comprehensive parameter monitor, FFU central controller, etc.
Air clean equipment:
Clean bench, clean shower, FFU, biology safety cabinet, clean sampling vehicle, clean work station, clean hand dryer, clean ventilation unit, heat exchange reclaim facility.
Air clean equipment to remove static electricity:
Clean bench, clean shower, clean work station to eliminate the static electricity and other ion blower fan, blower gun, blower fan, etc and other static electricity removal products.
We are striving to systematically expand our position as an innovative, customer-oriented technology corperation. Our objective is to create lasting value for our customers.