Product Specifications
Prod Model: |
5E-MAG6700 |
Markets: |
North America,South America,Eastern Europe,Southeast Asia,Africa,Oceania,Mid East,Eastern Asia,Western Europe |
Product Description
Quick Details
Place of Origin: Hunan, China (Mainland)
Brand Name: 5E
Model Number: 5E-MAG6700
Name: 5E-MAG6700 Automatic Proximate Analyzer
Comforms to Method: ASTM D7582, ISO7246 and GB/T212
Max.Sample loading: 19
Packaging & Delivery
Packaging Details: | Standard Export Package. Suitable for long distance delivery. |
---|
Delivery Detail: | Depend on stock, feel free to contact us. |
---|
MAG6700
5E-MAG6700 Automatic Proximate Analyzer is used to determine the moisture, ash, volatile matter and fixed carbon in coal and coke. It can also determine residual ash and slag combustibles in accordance with ASTM, which is widely applied in power plant, coal mine, metallurgy, chemical industry, commercial inspection, scientific research, education, etc.
Features
High Efficiency
Unattended operation, operator involvement is limited to just adding sample to a crucible.
High Stability and Reliability
1. Dual furnace with unique Volatile Matter furnace design for accurate result of volatile matter. It could work both simultaneously and independently.
2. Adopting advanced DSP controlling technology and updated adaptive control system.
3. Round shape furnace has solved conventional problem of unstable and uneven temperature.
4. Patented technology of heat-resistant to ensure stability and reliability.
5. Equipped with reliable sample sending system.
6.Adopting Thermogravimetric technology, using blank crucible to calibrate in the same atmosphere to ensure the precise weighing.
Convenient Operation
1. More analysis methods for choice, classical method and quick method.
2. Various samples weighing method. User has choice of sample loading continuously and sample loading singly.
3. Real time display of balance reading, which makes it convenient to control the precision of sample weight.
4. Query, print and calculate the experimental data during analysis is available.
5. Thermogravimetric technology for weight calibration at higher temperature, record the weight straightly.
Cost Saving
Available for analysis without oxygen and nitrogen by configured air pump, except for lignite.
Good Safety
Automatically heating off when open the ash furnace cover.
MODEL | 5E-MAG6700 |
Comforms to Method | ASTM D7582, ISO7246 and GB/T212 |
Max.Sample loading | 19 |
Furnace | Dual furnace |
Analysis Time | ≤120min for 19 samples |
Sample Mass | 0.8~1.2g recommended/ up to 5g |
Temp.Range | Up to 1050°C |
Analysis Precision | As per ASTM D5142-09 |
Temp.Controlling Resolution | ±5°C |
Power | FornaceI ≤4kW, FornaceII ≤5kW, |
Net weight | FurnaceI:80kg,Furnace II:50kg |
Dimensions(L×W×H) | Furnace I:550mm×580mm×890mm Furnace II:550mm×580mm×530mm |